News

Designing Endoscope Fiber Bundle Illumination in TracePro

Written by Admin | Aug 21, 2026, 1:00:00 PM

An endoscope that lights the tissue brightly at the center and leaves the field edges in shadow hands the surgeon an image where the periphery is unreadable. The illumination path in an endoscope is a fiber bundle that rings the imaging channel, carrying light from an external source to the distal tip, and three things routinely go wrong with it: the source couples poorly into the bundle and wastes flux, the numerical aperture of the delivered light does not match the field the camera sees, and the ring geometry of the illumination fibers casts a dark center or a bright halo on the tissue plane.

The design problem is to deliver uniform irradiance across the imaged field at the working distance, with enough total flux for the camera, through a fiber bundle whose numerical aperture and distal geometry the designer controls. TracePro models the endoscope illumination path as a physical system: the source coupling into the bundle, the guided light through fibers with a real numerical aperture, and the distal tip that launches light onto the tissue. Tracing that path with Monte Carlo ray tracing produces the irradiance map on the tissue plane, so uniformity and coupling efficiency are numbers rather than bench surprises. This article covers endoscope illumination design in TracePro.

The Illumination Problem in Endoscope Design

Endoscope illumination has to satisfy several requirements at once, and a design that optimizes one at the expense of the others produces an image the clinician cannot use.

Uniformity across the imaged field is the first requirement. The camera sees a field set by its own optics, often a wide field with a half-angle beyond 40 degrees, and the illumination must cover that field evenly at the working distance. A working distance for an endoscope is short, commonly 10 to 50 millimeters, and uniformity at that near range is harder than in the far field because the discrete illumination fibers have not yet blended. A ring of illumination fibers around a central imaging channel tends to overlap on axis and fall off at the field edge, or in a tightly collimated launch, to leave a dark spot on axis where no fiber points.

Sufficient flux is the second requirement. The camera needs a working signal level, and every coupling and guiding loss between the source and the tissue reduces it. Numerical aperture match is the third. Light launched from the distal tip into a cone wider than the camera field is wasted outside the image, and light launched into a cone narrower than the field leaves the edges dark. Matching the delivered numerical aperture to the camera field is what makes the flux land where the image needs it. TracePro handles all three because it traces the real fibers, the real launch geometry, and the real field.

Coupling the Source into the Bundle

The first loss in an endoscope illumination path is at the entrance to the fiber bundle, and it is governed by etendue and numerical aperture. A design that ignores this coupling throws away flux the rest of the system then struggles to recover. A fiber bundle accepts light only within its acceptance cone, set by its numerical aperture, and only across its active fiber area. A source whose etendue exceeds the product of the bundle area and its acceptance solid angle cannot fully couple, and the excess flux is lost at the input face no matter how bright the source. An LED or a short-arc lamp focused onto the bundle input with a condenser has to match its focused spot size and cone angle to the bundle area and acceptance angle. TracePro traces the source and condenser onto the bundle input face and reports the fraction of source flux that enters within the acceptance cone, which is the input coupling efficiency.

Matching numerical aperture at the input

Illumination fiber for endoscopes typically carries a numerical aperture in the range of 0.5 to 0.66, which corresponds to an acceptance half-angle from about 30 to 41 degrees. Focusing the source into a cone wider than the fiber acceptance overfills the numerical aperture, and the overfilled rays either fail to guide or leak out along the bundle. TracePro shows overfill directly, because rays entering beyond the acceptance angle are not guided, so the model reveals when a faster condenser is wasting flux against the fiber limit.

Modeling the Fiber Bundle and Distal Tip

The bundle carries the coupled light to the distal tip, and the tip launches it onto the tissue. TracePro models both stages so the delivered field reflects the real fiber properties and the real tip geometry rather than an idealized point of light. The bundle is modeled with its numerical aperture, which sets both the acceptance cone at the input and the emission cone at the output, since a step-index fiber preserves the numerical aperture end to end. A tightly bent bundle can leak guided rays where a bend drops the internal angle below the guiding limit, and TracePro traces the bend geometry so any bend-induced loss enters the delivered flux. The individual illumination fibers can be modeled as a group, arranged in the real ring pattern around the imaging channel, so the spatial distribution of the launch points is correct.

The distal tip is where uniformity is set. A bare fiber ring launches light in the numerical-aperture cone of the fiber, and whether that produces a uniform field depends on the ring diameter, the fiber numerical aperture, and the working distance. Adding a diffusing window, a molded lens, or an angled polish at the tip reshapes the launch, and TracePro traces each option to show its effect on the tissue-plane field. A diffuser at the tip widens and smooths the launch to fill a dark center, while a tilted launch aims the ring outward to reach the field edges.

Irradiance Uniformity on the Tissue Plane

The design is judged on the tissue plane at the working distance, and TracePro computes the irradiance map there. The map shows the delivered field the way the camera would see it lit, so a dark center, a bright ring, or an edge falloff appears as data rather than as a rendered impression.

Evaluating the map across the camera field, from the axis to the field edge at the working distance, gives the uniformity the clinician experiences. A center-dark map from a collimated fiber ring calls for a tip diffuser or a wider launch cone. A center-bright map from an overlapping launch calls for a larger ring diameter or a narrower cone that pushes flux outward. Because the map is computed at the real short working distance, it captures the near-field behavior that a far-field approximation would miss.

Uniformity has to hold across the working-distance range, not at a single plane. An endoscope is used at varying distances from tissue, so tracing the irradiance map at the near and far ends of the intended range shows whether the design stays acceptable as the tip moves. A launch tuned for uniformity at 20 millimeters that collapses to a hot center at 10 millimeters is a design that needs a broader tip treatment, and the model exposes that before clinical prototypes are built.

Balancing uniformity against delivered flux

Every step that improves uniformity, a stronger tip diffuser or a wider launch, spreads the same flux over a larger solid angle and lowers the peak irradiance. TracePro reports both the uniformity and the absolute irradiance at the tissue plane, so the design can be tuned to the point where the field is even enough for the image and still bright enough for the camera signal. Reading both numbers from one trace keeps the tradeoff explicit.

From Model to Clinical Prototype

Endoscope illumination design is a chain from source coupling, through a fiber bundle with a real numerical aperture, to a distal tip that sets the delivered field, and it is judged on tissue-plane uniformity at a short working distance. TracePro models the whole chain, reports coupling efficiency at the bundle input, and computes the irradiance map on the tissue plane so uniformity and flux are decided before hardware. That turns a design that would otherwise be tuned by building tips and looking at images into a set of traced choices verified against a uniformity target across the working-distance range.

Request a TracePro trial and model your endoscope illumination path, or contact Lambda Research to request a demo focused on fiber bundle illumination for medical devices.